Degaussing Instructions: Difference between revisions
(Created Degaussing Page) |
mNo edit summary |
||
Line 3: | Line 3: | ||
==Introduction== |
==Introduction== |
||
If a subject arrives with embedded metal on their person, i.e., a permanent retainer, which can not be removed it may be necessary to de-magnetize the subject.<br><br> |
If a subject arrives with embedded metal on their person, i.e., a permanent retainer, which can not be removed it may be necessary to de-magnetize the subject.<br><br> |
||
'''Care should be taken not to use the Degausser close to the computer monitors as it may permanently damage the screens |
'''Care should be taken not to use the Degausser close to the computer monitors as it may permanently damage the screens.'''<br> |
||
'''The Degausser must NOT be used in the MSR.''' </br> |
|||
==Directions== |
==Directions== |
||
*Plug in the Degausser. |
*Plug in the Degausser. |
Revision as of 10:24, 16 December 2021
Note: These are internal instructions for MEG Core Staff. If you are not MEG Core staff, you are welcome to educate yourself, but please do not attempt without supervision or direct instruction.
Introduction
If a subject arrives with embedded metal on their person, i.e., a permanent retainer, which can not be removed it may be necessary to de-magnetize the subject.
Care should be taken not to use the Degausser close to the computer monitors as it may permanently damage the screens.
The Degausser must NOT be used in the MSR.
Directions
- Plug in the Degausser.
- Move to an area* away from all computer monitors (minimum of 3-4 feet.)
- Approach the subject, passing the Degausser in a circular motion in front of the area of interest 3 or 4 times, and then back away repeating the process.
- Unplug Degausser.
*Note: The current layout of the MEG lab does not have sufficient space to complete this procedure without permanent damage to the surrounding electronics. The best alternative is to perform this procedure in the NMR courtyard or an empty hallway.
Additional Comments
- If used improperly you can magnetize your subject. If you are unsure about the degaussing procedure, ask for assistance from MEG staff before using the Degausser.
- Similarly, it is recommended that the subject not have had an MRI within 1-2 days before having a MEG scan because of the possibility of residual magnetization.