Troubleshooting: Difference between revisions

From MEG Core
Jump to navigation Jump to search
Content added Content deleted
Line 1: Line 1:
=== Procedure for restarting the rack ===
==== Procedure for restarting the rack ====


<span style="color: rgb(200, 0, 0);"><u>'''There are two important power buttons''' </u></span>
<span style="color: rgb(200, 0, 0);"><u>'''There are two important power buttons''' </u></span>
Line 24: Line 24:
::::::'''''*Please contact MEG staff for assistance if you are unable to resolve problem(s).'''''
::::::'''''*Please contact MEG staff for assistance if you are unable to resolve problem(s).'''''


==== RESETTING THE JOYSTICK /FORP CONTROLLER ====
==== Procedure for Resetting the JOYSTICK /FORP CONTROLLER ====


'''TO RESET THE JOYSTICK / FORP CONTROLLER:'''
'''TO RESET THE JOYSTICK / FORP CONTROLLER:'''


::If set to the joystick mode follow the instructions below to re-set to the FORP:
::If the Controller is set to the joystick mode follow the instructions below to re-set to the FORP:
::::1.Press the knob (shaft) on the right.
::::1.Press the knob (shaft) on the right.
::::2. The Display will read: “Change Modes?” Rotate to select and press the knob.
::::2. The Display will read: “Change Modes?” Rotate to select and press the knob.
Line 36: Line 36:
::::6. Select: “USB Key 12345” then press the knob.
::::6. Select: “USB Key 12345” then press the knob.
<div align="center">- DONE – </div>
<div align="center">- DONE – </div>

4/18/18

Revision as of 13:49, 19 April 2018

Procedure for restarting the rack

There are two important power buttons

The button on the DSQ 3500 Electronics Control Console (EC Console) with the ring of light controls the power to the electronics rack. This light changes colors when the rack is powered up.
The Head Localization Unit (HLU) power box, on the top of the new rack, has a button with a light within it. Press this button to turn on and off the HLU unit.


There are times that the MEG electronics rack needs to be rebooted.

  • If the Acq starts with some noisy channels, usually restarting the Acq will make the noise go away, if not reboot.
  • If the HLU does not work, you must reboot the entire rack.


To restart (reboot) the electronics rack:
  1. First you must make sure the you have closed ACQ.
  2. Then power off the rack. Press the button on the EC console.
  3. Next you must power off the HLU box which is located on top of the new rack. Press it to turn off the HLU.
  4. Return to the console and power the rack back on.
  5. Wait about 10 seconds.
  6. Next, turn the HLU box back on.
  7. Now you can launch Acq.
*Please contact MEG staff for assistance if you are unable to resolve problem(s).

Procedure for Resetting the JOYSTICK /FORP CONTROLLER

TO RESET THE JOYSTICK / FORP CONTROLLER:

If the Controller is set to the joystick mode follow the instructions below to re-set to the FORP:
1.Press the knob (shaft) on the right.
2. The Display will read: “Change Modes?” Rotate to select and press the knob.
3. Select: “Autoconfigure” (at the top). Press the knob.
4. Select: “Legacy Handheld” (a ways down). Press the knob.
5. Select: “USB” then press the knob.
6. Select: “USB Key 12345” then press the knob.
- DONE –

4/18/18