Lab Status: Difference between revisions

From MEG Core
Jump to navigation Jump to search
Content added Content deleted
No edit summary
 
(37 intermediate revisions by 5 users not shown)
Line 1: Line 1:
==Live cam of the MEG Core console (NIH access only)==
=Procedure for restarting the rack =


<div style="transform: scale(2); transform-origin: left 0 0">
<span style="color: rgb(200, 0, 0);"><u>'''There are two important power buttons''' </u></span>
https://kurage.nimh.nih.gov/nih/megcam.jpg
: The button on the console with the ring of light controls the power to the rack. This light changes colors when the rack is powered up.
<br>
: The HLU has a button with a light within it. Press this button to turn on and off the HLU unit
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
</div>
{| width=512 style="border: 1px solid grey;"
| Image updated every 5 minutes.
|}


;'''There are times that the MEG rack needs to be rebooted.'''
'''The main MEG lab phone number is (301)402-2445.'''
* If the Acq starts with some noisy channels, usually restarting the Acq will make the noise go away.
* If the HLU does not work, you must reboot the entire rack.

To restart (reboot) the rack:

# First you power off the rack. Press the button on the console.
# Next you must power off the HLU box which is located on top of the new rack. Press it to turn off the HLU.
# Return to the console and power the rack back on.
# Wait about 10 seconds.
# Next, turn the HLU box back on.
# Now you can launch Acq.

Latest revision as of 13:04, 15 August 2023

Live cam of the MEG Core console (NIH access only)

megcam.jpg











Image updated every 5 minutes.

The main MEG lab phone number is (301)402-2445.