Quality Assurance
MEG Calibration Procedures
Calibration procedures should be performed for quality assurance and in order to evaluate the performance of the MEG system.
CALIBRATION PROCEDURE: PERFORM WHENEVER: 1. Noise / Gain measurement line frequency noise has appeared in the noise collection measurement 2. Head coil calibration head localization results / accuracy is in doubt 3. Magnetic phantom measurement the performance of the MEG system is in question 4. EEG calibration EEG results are uncertain